This product is no longer available.
Please check new generation in the section predecessor / successor or contact your local Endress+Hauser entity.
Level measurement, capacitive.
Application: liquids.
Rod probe.
Suitable for small vessels +
heavy process conditions.
:: High pressure / high temperature
:: Suitable at heavy stress from side
:: Longtime stable tightness
-
Successor
-
Standard products
- Reliable, robust and low-maintenance
Technical excellence
Simplicity
-
Specialized products
- Designed for demanding applications
Technical excellence
Simplicity
Variable
FLEX selections Technical excellence Simplicity - F
- L
- E
- X
Fundamental selection
Meet your basic measurement needs
Technical excellence Simplicity - F
- L
- E
- X
Lean selection
Handle your core processes easily
Technical excellence Simplicity - F
- L
- E
- X
Extended selection
Optimize your processes with innovative technologies
Technical excellence Simplicity - F
- L
- E
- X
Xpert selection
Master your most challenging applications
Technical excellence Simplicity Variable
-
-
Characteristic / Application
Fully/partly insulated rod probe with 3/4" Processconnection, suitable for heavy process condition
T: Standard application
E: European Certificate (EEx d, EEx ia) -
Specialities
Interface measurement, Foam detection
-
Supply / Communication
3-wire
PFM -
Ambient temperature
Standard:
-40 °C ... 70 °C
With separate housing:
-50°C ... 120 °C -
Process temperature
-80 °C ... 200 °C
-
Process pressure / max. overpressure limit
Vacuum ... 25 bar
-
Main wetted parts
Insulation material: PTFE, PFA, PE
316L -
Process connection
G 3/4A, G 1A /NPT 3/4", NPT 1"
Flange from DN25.../ASME 1".../JIS... -
Process connection hygienic
Tri-Clamp ISO2852
Dairy coupling -
Sensor length
Total length of probe 3 m
-
Communication
3-wire
PFM -
Certificates / Approvals
ATEX
WHG -
Options
Separate housing (HTC)
-
Components
Transmitter
3-wire: FTC325
PFM: FTC325, FTC625, FTC470Z, FTC471Z -
Successor
FTI51
-
Application limits
Notice the pressure and temperature derating.
Notice for non conductive medium the DK.
-
Characteristic / Application
Fully/partly insulated rod probe with 3/4" Processconnection, suitable for heavy process condition
T: Standard application
E: European Certificate (EEx d, EEx ia) -
Specialities
Interface measurement, Foam detection
-
Supply / Communication
HART
PROFIBUS PA
3-wire
PFM -
Linearity error for conductive liquids
< 1% at 1m
-
Ambient temperature
Standard:
-40 °C ... 70 °C
With separate housing:
-50°C ... 120 °C -
Process temperature
-80 °C ... 200 °C
-
Process pressure / max. overpressure limit
Vacuum ... 25 bar
-
Main wetted parts
Insulation material: PTFE, PFA, PE
316L -
Process connection
G 3/4A, G 1A /NPT 3/4", NPT 1"
Flange ab DN25.../ASME 1".../JIS... -
Process connection hygienic
Tri-Clamp ISO2852
Dairy coupling -
Sensor length
Total length of probe 3 m
-
Max. measurement distance
0.1 m ... 3.0 m
-
Communication
HART
PROFIBUS PA
3-wire
PFM -
Certificates / Approvals
ATEX
WHG -
Options
Separate housing (HTC)
-
Components
Transmitter
3-wire: FMC420
PFM: FMX570, FMC671Z, FMC661 -
Successor
FMI51
-
Application limits
Notice the pressure and temperature derating.
Notice for non conductive medium the DK.